Original Article
Comparison of Central Corneal
Thickness Measurement Using Non-Contact and Contact Pachymetry
Devices in Normal Eyes
Qamar-ul-Islam, Sidra
Malik
Pak J Ophthalmol
2015, Vol. 31 No. 1
. . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . .. . .. . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . .
|
See end of article for authors affiliations …..……………………….. Correspondence to: Surg Cdr Qamar Ul Islam Classified
Eye Spec /Asst Prof PNS
Shifa /Bahria University
Med & Dental College (BUMDC) Karachi …..……………………….. |
Purpose: To compare central corneal thickness (CCT) in normal population
using contact and non-contact pachymetry devices
and to assess the intra operator repeatability of measurement with each
device. Materials and Methods: This prospective, cross sectional comparative study evaluated 30
healthy subjects fulfilling the inclusion criteria reporting in AFIO
Rawalpindi. CCT was measured in both eyes of subjects using noncontact
specular microscope, Dual Scheimpflug analyzer and
contact ultrasound pachymeter by a single
investigator at the same time of the day. Pearson’s correlation coefficient
test was performed to ascertain correlation between pachymetry
devices. Intra operator repeatability was analysed
using within subjects coefficient of variation/repeatability (CoV) and intra class correlation coefficient (ICC). Results:
Sixty eyes of 30 male subjects were analysed.
Mean age of study population was 31.03 ± 10.30 years. Mean CCT values were
536.48 ± 35.77 µm, 498.62 ± 34.70 µm and 526 ± 37 µm with Dual Scheimpflug Analyzer , Specular microscope and Ultrasonic
pachymeter respectively (p < 0.01). There was
significant linear co relation between all measurement modalities (r = 0.804
to r = 0.949) (p < 0.01). Intra operator repeatability was excellent for
all devices as indicated by low CoV values (<
0.80%) and high ICC values (> 0.90). Conclusion:
All three devices showed excellent intra operator repeatability
for CCT measurement making them reliable tools for CCT measurement. Key Words: Central Corneal Thickness, Pachymetry, Repeatability, Microscopy. |
Corneal thickness is a sensitive
indicator of corneal hydration and patency of corneal endothelial pump.
Accurate central corneal thickness (CCT) measurement (Pachymetry)
has diagnostic and therapeutic implications in various conditions like ectatic corneal dystrophies (Keratoconus,
Pellucid marginal degeneration), contact lens related problems, dry eyes,
diabetes mellitus, glaucoma and refractive surgery (LASIK).1,2 An
ideal method of pachymetry should be accurate, safe,
repeatable, reproducible, easy and quick to perform.3 Over a period
of time, many methods have been devised for pachymetry
such as ultrasound pachymetry, ultrasound bio
microscopy, slit scanning corneal topography, confocal microscopy, optical
biometry, Scheimpflug system, specular microscopy,
spectral domain OCT and very high frequency ultrasound scanner (VHFUS).2,3
For years, ultrasound pachymetry remains the gold standard method for measurement
of CCT due to its high degree of intra operator and inter operator
reproducibility.4 But the potential disadvantages of this method
include possibility of probe malplacement and malalignment, inadvertent indentation leading to slightly
thinner readings, patient’s discomfort, need for topical anaesthesia,
epithelial damage and risk of infection.1,4-8 In the current era,
newer non-invasive methods of pachymetry have been
thoroughly evaluated for accuracy, precision, repeatability, reproducibility,
and agreement between these new devices and gold standard ultrasound pachymeter. Many studies have demonstrated acceptable
repeatability and reproducibility of various non-contact pachymetry
equipments and conformed agreement between CCT measurements with non-contact
methods and contact ultrasound methods.1,4, 7,9,10
The objective of this study was to compare CCT measurements
in healthy volunteers using contact ultrasound pachymeter,
noncontact Dual Scheimpflug analyzer and non-contact
specular microscope and to assess the intra operator repeatability of
measurement with each device.
MATERIAL AND METHODS
This prospective, cross sectional
comparative study enrolled 30 consecutive healthy subjects (Staff members,
candidates for medical examination and attendants of patients) reporting in
Armed Forces Institute of Ophthalmology (AFIO) Rawalpindi between 21 June 2013
to 30 Aug 2013. Subjects with ocular or systemic disease, history of ocular
surgery or trauma, intraocular pressure (IOP) > 21 mm Hg, refractive error
≥ ± 1.5 dioptres, contact lens wearers and
those using any topical ocular medications were excluded. The study was conducted
in accordance with the ethical considerations given in Helsinki declaration and
written and informed consent was obtained from each subject before examination.
Comprehensive ophthalmic examination including visual acuity, refraction, slit
lamp examination and fundus examination was carried out in each subject.
CCT was measured in both eyes of
subjects using noncontact specular microscope (SP 3000 P; Topcon, Japan), Dual Scheimpflug analyzer (Galilei™
G4; Ziemer, Switzerland) and contact ultrasound Pachymeter (IOPac® Advanced;
Reichert). All the readings were taken by a single investigator at the same
time of the day between 1000 – 1400 hours to avoid diurnal variation of pachymetry readings. For each subject all the measurements
were performed within a 45 minute period. CCT readings were first taken by
non-contact method (Specular microscope or Dual Scheimpflug
Analyzer). Ten readings were recorded for each eye with a gap of 30 seconds
after each reading and fresh alignment of equipment was done each time.
Following measurements with both non-contact equipments, cornea was
anaesthetized with topical 0.5% Proparacaine
Hydrochloride (Alcaine) eye drops and 10 readings on
each eye were taken with ultrasound pachymeter by
placing the sterile ultrasound probe perpendicular to the centre of cornea and
asking the patient to blink before each new reading. All the measurements were
endorsed on a pre-devised proforma.
Statistical analysis of the data was
done using SPSS version 13.0. All the data were tested for normality before
analysis. Descriptive statistics i.e. means ± standard deviation (SD) for
quantitative variables and frequencies and percentages for quail-tative variables were used. Inter device differences were analysed using paired sample ‘t’ test.
Pearson’s correlation coefficient test was performed to compare the mean CCT
values obtained from non-contact equipment with contact ultrasound pachymeter. A p value of ≤ 0.05 was considered
significant. Intra operator repeatability was analysed
using within subjects coefficient of variation / repeatability (CoV) and intra class correlation coefficient (ICC). CoV was defined as the SD values divided by the mean
result. The lower the CoV and higher the ICC the more
repeatable the measurements were. For repeatability assessment 10 readings of
right eye of 10 randomly selected subjects were used.
RESULTS
Sixty eyes of 30 healthy volunteer male
subjects fulfilling the inclusion criteria were analysed.
Mean age of study population was 31.03 ± 10.30 years (Range: 16- 50 Years) with
33.33% of subjects were in 3rd decade of life. Mean CCT values were
536.48 ± 35.77 µm , 498.62 ± 34.70 µm and 526 ± 37 µm with Galilei™
G4 Dual Scheimpflug Analyzer , SP 3000 Specular
microscope and IOPac Advanced Ultrasonic pachymeter respectively (p < 0.01). Mean CCT values for
right and left eyes were comparable for each measurement modality (p > 0.05)
(Table 1). Inter device comparison results showed that mean CCT values obtained
by specular microscope and ultrasound pachymeter were
7.05% and 1.90% lower than Galilei analyzer values,
while CCT measurement obtained by specular microscope were 5.25% lower than
those from ultrasound method (p < 0.01). There was strong linear co relation
between all measurement modalities with Pearson co relation coefficient ranging
from r = 0.804 to r = 0.949 (p <0.01) (table 2).

The coefficient of variation (CoV) and Intra class correlation coefficient (ICC) for CCT
measurement from different methods is given in table 3. Agreement for
successive measurements performed was excellent for all devices as indicated by
low CoV values (< 0.80%) and high ICC values (>
0.90). Galilei Dual Scheimpflug
analyzer produced lowest CoV (0.406%) and Specular
microscope gave highest ICC values (0.996) indicating a high degree of intra
operator repeatability of these equipments.
DISCUSSION
Importance of CCT measurement in
various domains of ophthalmology including diagnosis of glaucoma, cataract and
refractive surgery cannot be undermined. Availability of a precise and accurate
non-contact pachymetry device with high intra
operator repeatability eliminates the need of conventional ultrasonic pachymeter with its potential contact hazards. Most of the
studies on CCT measurements in Pakistan have been done using various contact
ultrasound pachymeter. Mean CCT values in our study
were 526.27 ± 37 µm using ultrasonic pachymeter that
were comparable to CCT values of 535.68, 531.08 and 540.60 µm quoted in various
studies on Pakistani population.11-13 In our study, mean CCT values
using Galilei Dual Scheimpflug
analyzer, SP- 3000 Specular microscope and Ultrasonic pachymeter
were 536.48 ± 35.77 µm , 498.62 ± 34.70 µm and 526 ±
37 µm respectively. A lot of work has been published in international
literature regarding CCT measurement by different devices in various ethnic and
racial groups (table 4). Most of the results from these studies are comparable
to our results, whereas, observed differences from our results may possibly be
due to racial differences, variation in study settings and measurement methods.
Overall, inter device comparison in our study showed mean paired differences of
CCT values being statistically significant between devices (p < 0.01).
However, pairwise comparison of all devices demonstrated significantly strong
linear correlation with Pearson correlation coefficient ranged from r = 0.804
to r =0.949. Study by Ou TH et al, showed significant
good linear correlation between ultrasound pachymeter
- Orbscan II (r = 0.793, p < 0.001) and ultrasound
pachymeter – SP 3000 specular microscope (r = 0.890,
p < 0.001) for CCT measurement in eyes with glaucoma or glaucoma suspect.5

Over a period of last decade, various
computerized corneal tomography devices were developed based on the principle
of placido disc and slit scanning (Orbscan II), single rotating Scheimpflug
camera (Pentacam) and dual Scheimpflug
and placido disc (Galilei)
that provide more reliable pachymetry data apart from
providing accurate curvature and topographic analysis of cornea. In a study by
Crawford AZ et al Orbscan II measured significantly
lower CCT values (524 ± 36 µm) compared with Galilei
(542 ± 26 µm) and Pentacam (544 ± 26 µm).18
However, CCT values obtained by Galilei dual Scheimpflug analyzer were comparable with our results. In
our study, CCT values obtained by specular microscope were significantly lower
than both ultrasound pachymeter and Galilei analyzer (p = <0.01). But this pattern of
significantly lower CCT measurements obtained by specular microscopy as
compared to other modalities had been reported in other studies as well.1,5,14,15
The difference between specular microscope pachymetry
and other two methods used in our study was probably due to different operating
principles. The non-contact specular microscopy is based on reflection of
light, while ultrasonic pachymetry depends on
reflection of ultrasonic waves from anterior and posterior corneal surfaces and
Galilei dual Scheimpflug
analyzer uses two rotating Scheimpflug cameras 180
apart along with placido imaging.
Comparing repeatability of the
instruments is essential because it reflects amount of agreement that is
possible between instruments. We used coefficient of variation (CoV) and intra class correlation coefficient (ICC) to
ascertain intra operator repeatability and our results suggested excellent
agreement of repeated measurements for all devices as indicated by low CoV values (< 0.80%) and high ICC values (> 0.990).
In our study Galilei analyzer produced lowest CoV (0.40%) and specular microscope gave highest ICC
(0.996) indicating high degree of repeatability of these instruments. Weerawat K et al, reported high intra operator
repeatability (ICC = 0.985) for CCT measurement by ultrasonic pachymeter that was comparable to our repeatability results
using ultrasonic pachymeter (ICC = 0.995).19
Various studies demonstrated high intra operator repeatability of various
non-contact pachymetry devices (Galilei,
Orbscan, specular microscope and SD OCT) and
ultrasonic pachymeter for CCT measurement as
reflected by low CoV (between 0.33 – 0.93%) and high
ICC values (> 0.978).3-4,10,15,18
Discrepancies between studies may be
due to several factors, including statistical method, subject characteristics,
and the definition of acceptable agreement. Subject characteristics also may
influence the interpretation of agreement. One of the limitations of our study
was that we examined both eyes of each subject. Although this increases the
sample size, it raises the fundamental issue of inter eye correlation. The use of both eyes of each subject
therefore may account for some of the differences reported between studies. In the current study, the pachymetry values obtained by the Galilei,
Specular microscope, and ultrasonic pachymeter were
sufficiently disparate that the 3 devices could not be considered equivalent.
As the true gold standard for CCT measurement is not yet established, it is
difficult to conclude which device obtains the most accurate measurements, and
therefore it is not possible to recommend one device absolutely over the others
at the present time.
CONCLUSION
Both non-contact devices showed excellent
intra operator repeatability for CCT measurement that was comparable to contact
method in normal eyes making them reliable tools for CCT measurement providing
better patient comfort. Although, all three devices showed strong linear
correlation , they are not inter changeable
as mean CCT measurement values differ significantly between instruments.
Author’s Affiliation
Dr.
Qamar Ul Islam
Classified
Eye Spec /Assistant Professor
PNS Shifa /Bahria University Medical and Dental College (BUMDC), Karachi
Dr. Sidra Malik
AFIO
Rawalpindi
REFERENCES
1.
Almubrad TM, Osuagwu UL, AlAbbadi I, Ogbuehi KC. Comparison
of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clinical Ophthalmology. 2011; 5: 871–6.
2.
Al-Farhan HM,
Al-Otaibi WA. Comparison of central corneal thickness measurements using
ultrasound pachymetry, ultrasound biomicroscopy,
and the Artemis-2 VHF scanner in normal eyes. Clinical Ophthalmology. 2012; 6:
1037–43.
3.
Piotrowiak
I, Soldanska B, Burduk M, Kaluzny BJ, Kaluzny J. Measuring Corneal Thickness with SOCT,
the Scheimpflug System, and Ultrasound Pachymetry. ISRN Ophthalmology 2012, Article ID
869319, 5 pages.doi:10.5402/2012/869319.
4.
Ladi
SJ, Shah AN.
Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer
and ultrasound pachymetry. Indian J Ophthalmol. 2010; 58: 385–8.
5.
Ou
TH, Lai IC, Teng MC. Comparison of Central Corneal
Thickness Measurements by Ultrasonic Pachymetry, Orbscan II, and SP3000P in Eyes with Glaucoma or Glaucoma Suspect.
Chang Gung Med J. 2012; 35: 255-62.
6.
Guilbert
E, Saad A, Grise-Dulac A, Gatinel D. Corneal thickness, curvature, and elevation readings in
normal corneas: Combined Placido–Scheimpflug
system versus combined Placido–scanning-slit system.
J Cataract Refract Surg. 2012; 38: 1198–1206.
7.
Shafiq
I, Hashmani S. Comparison Between Central Corneal
Thickness Measurements Obtained with Orbscan II
Topographer and Ultrasonic Pachymeter. Pak J Ophthalmol. 2007; 23: 87-90.
8.
AlFarhan
HM. Measurements of
central corneal thickness using two immersion ultrasound techniques and optical
technique. JPMA. 2014; 64: 266-70.
9.
Yekta
AA, Hashemi H, Khabaz Khoob M, Dostdar A, Mehravaran S, Heravian J, et al. Comparison of Central Corneal
Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter.
Iranian J of Ophthalmol. 2009; 21: 51-7.
10.
Bayhan
HA, Bayhan SA, Can I. Comparison of central corneal
thickness measurements with three new optical devices and a standard ultrasonic
pachymeter. Intl J Ophthalmol.
2014; 7: 302-8.
11.
Ahmed J, Memon
MF. Central Corneal
Thickness and its Relationship with Myopia. JLUMHS. 2008; Jan-Apr: 4-6.
12.
Channa
R, Mir F, Shah MN, Ali A, Ahmad K. Central corneal thickness of Pakistani adults. JPMA.
2009; 59: 225-9.
13.
Akram
S, Anklesaria ZH, Ahmad K. Correlation between Central Corneal
Thickness Measurements Using Two Different Ultrasonic Pachymeters.
Pak J Ophthalmol. 2013; 29: 214-6.
14.
Ogbuehi
KC, Osuagwu UL. Repeatability and interobserver
reproducibility of Artemis – 2 high – frequency ultrasound in determination of
human corneal thickness. Clinical Ophthalmology. 2012; 6: 761-9.
15.
Bao
F, Wong Q, Chong S, Savini G, Lu W, Feng Y, et al. Comparison and evaluation of central corneal thickness using
2 new noncontact specular microscopes and conventional pachymetry
devices. Cornea. 2014; 33: 576-81.
16.
Wong AC-M, Wong CC, Yuen NSY, Hui SP.
Correlational study of central corneal thickness measurements on Hong Kong
Chinese using optical coherence tomography, Orbscan
and ultrasound pachymetry. Eye 2002; 16: 715–21.
17.
Faramarzi
A, Karimian F, Jafarinasab
MR, Jabbarpoor Bonyadi MH, Yaseri M. Central corneal thickness measurements after myopic
photorefractive keratectomy using Scheimpflug
imaging, scanning-slit topography, and ultrasonic pachymetry.
J Cataract Refract Surg. 2010; 36: 1543–9.
18.
Crawford AZ, Patel DV, Mcghee CNJ. Comparison and Repeatability of Keratometric
and Corneal Power Measurements Obtained by Orbscan
II, Pentacam, and Galilei
Corneal Tomography Systems. Am J Ophthalmol. 2013;
156: 53–60.
19.
Kiddee
W, Horattanareung O. Intraoperator
Repeatability and Interoperator Reproducibility in
the Ultrasonic Pachymetry Measurements of Central
Corneal Thickness. J Med Assoc Thai. 2009; 92: 672-6.